Main features:
The automatic control system can not only detect the surrounding cells automatically, but also adjust the pressure control micro thruster to automatically detect the increase or decrease of electrode resistance of cell electrode, automatically detect neurons and form high resistance sealing. G ohm sealing, membrane breaking and recording software automatically calculate and automatically control without human intervention.
When the electrode enters the experiment, the system can control and automatically calculate the positive pressure output accurately. After the cell is formed with high resistance sealing, the system can automatically calculate and accurately control the negative pressure output.
The system can sequence single cells and experiment rats under awake and anaesthesia.
The system can also be used in the experiment of tissue slice in vitro. The software control is simple, convenient and flexible, and has a wide scalability.
Operation:
The control system can be used with the software developed by our company. It will issue appropriate serial commands to the control unit during operation to obtain the whole cell record. These commands are listed here for reference or part 3 development.
The software also controls the motor position connected to the probe, which works in collaboration with Newport pzc200 KT.
The computer control pressure system in the control unit can switch between the patch clamp and the four discrete pressures of each stage of the following serial command:
Pressure range (mbar) command
Atmosphere 0 switchatom
Low pressure 15-344 switchlow
High pressure 50-1000 switchhighp
Low 7.5-150 switchlow
7.5-150 switch at high altitude
All pressures can also be changed by knob in unit or using software serial port command panel.
USB cable to PC
COM port will appear in Windows Device Manager to connect to the computer via USB.
Effective value of pressure command
Low pressure setlowp, 15-344
High pressure sethighp, 50-1000
Low altitude setlowv, 7.5-150
High altitude sethighv, 7.5-150
The representative of #, is ordered to accept the number
For example, set pressure command:
setlowP,025
The minimum output negative pressure of the above command is 25mbar
setlowP,025
technical specifications
Maximum input pressure 150psi
Minimum input pressure 50psi
Power supply requirement 24 vdc/3a
High pressure output range 0-20psi
Low pressure output range 0-5psi
Low / high vacuum output range 0-14psi
Transmission speed USB 2.0
Compatible operating system win7
The control system also has breakthrough command, which can apply high negative pressure to electrode (MS) according to the interval time specified by the user
For example, set the break through command:
breakIn,300
“Breakin” is the instruction, 300 is the length of time, in milliseconds, and high negative pressure will act on the electrode.
There is also an embedded relay control unit, switch, analog output for the internal command connection of patch clamp amplifier, and the internal data and external input signals between the analog output of the switch are connected to the analog BNC connected to the front panel. Here is the command to switch the relay.
Sample relay switch command:
BNCswitch,ON
The two valid parameters of this command are on or off
These are general settings for serial communication
Serial port setting
Baud rate 9600
Data bit 8
Parity either side
Stop bit 1
Model | Price |
Automatic patch-clamp system | €75000 |